3M™ Comply™ Bowie-Dick Type Test Systems 1233LF, 30 per case
Couldn't load pickup availability
3M 1233LF
Use daily on every pre/post vacuum autoclave (T-Edge 11B, 10B, Elara, Bravo 17 and 21, StatClave, Steris 3013, 3023, 3033, 3043, 3053)
Features:
- Easy-to-read test sheet clearly identifies failures and can be filed for record keeping
- Lead Free Ink Chemistry
- Equivalent to the Bowie-Dick Towel Pack cited in AAMI.
- Ink is not made with lead materials
- 3M™ Comply™ Bowie-Dick Plus Test Pack 1233LF is equivalent in performance to the Bowie-Dick Towel Pack described in ANSI/AAMI ST79
- Two year shelf life
- Designed for testing air-removal efficiency of 132-134˚C (270-273˚F) dynamic-air removal steam sterilizers
- Test pack will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum by testing the air removal efficiency of 132-134˚C (270-273˚F) Dynamic Air-Removal steam sterilizers
- Indicator sheet will turn a uniform dark brown/black color except when air removal failures, such as air leaks, occur
As an Equipment Monitor, the 3M™ Comply™ Bowie-Dick Test Pack 1233LF is designed for testing air removal efficiency of 132-134˚C (270-273˚F) dynamic-air-removal-steam sterilizers. Test packs will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum. Complies to ANSI/AAMI/ISO 11140-5:2007 and are Type 2 (Category s2) Special Indicators as categorized by ISO 11140-1:2014.
Suggested Applications
A Bowie-Dick test should be run daily on each sterilizer. Should also be used to check sterilizer performance after a major repair.
- Overall Height: 5.118 in (13 cm)
- Overall Length: 4.331 in (11 cm)
- Overall Width: 0.787 in (2 cm)
- Size: 0.787 in x 4.331 in
- Size: 2 cm x 11 cm
- Units per Case 30
Discover more in our FAQ
Shoot our team an email & we’ll get back to you ASAP
How do I look up disinfectant on List N?
How do I look up disinfectant on List N?
First, find the product registration number. This can be found on the product label or safety data sheet. Duraline posts most registrations in the Product Description field of the product page in our website. Second, go to the epa.gov website: https://cfpub.epa.gov/giwiz/disinfectants/index.c... and enter the first 2 parts of the registration number and click enter.
My product has 3 parts on the registration number and I can't find all 3 on the approved website for List N. Why?
My product has 3 parts on the registration number and I can't find all 3 on the approved website for List N. Why?
The first 2 parts of the EPA Registration Number are the primary numbers and represent the primary registration. Since different companies sell products with the same chemical composition and efficiency, only 1 brand or product name receives the primary number. The third part represents the distributor of the primary registration and is still on List N and is proven effective against SARS-CoV-2 (COVID-19).
I can't find Swovo disinfectant wipes on List N. Where is it?
I can't find Swovo disinfectant wipes on List N. Where is it?
The EPA Registration number for Swovo is 59894-10-98425
To find this on the epa.gov website, you only need to enter the first 2 parts of this number, "59894-10"
This primary registration number is on the list, and so it is approved! See screenshot below.