Option 1: Roof Venting Install with a 2" SCH 40 PVC pipe and roof vent boot. Andersen offers a ¾" I.D. to 2" adapter cap for secure connection.
Option 2: Remote Location Place the sterilizer up to 200 feet (60.96 m) away using ¾" braided tubing. Choose a protected path to prevent obstruction or damage.
Option 3: Wall, Soffit, or Eve Termination The through-wall mounting kit can also be used for ceiling mounts to vent through soffits or eves. Vent screens and straight quick disconnect fittings are available.
Important Compliance Note: All EtO sterilizers must be vented through a dedicated exhaust line directly outdoors or through an emission control system. Venting should not terminate within 25 feet (7.6 m) of building air intakes, pedestrian areas, or animal traffic.